What: /sys/bus/iio/devices/iio:deviceX/calibration_auto_enable
|
|
Date: June 2020
|
|
KernelVersion: 5.8
|
|
Contact: linux-iio@vger.kernel.org
|
|
Description:
|
|
Contaminants build-up in the measurement chamber or optical
|
|
elements deterioration leads to sensor drift.
|
|
|
|
One can compensate for sensor drift by using automatic self
|
|
calibration procedure (asc).
|
|
|
|
Writing 1 or 0 to this attribute will respectively activate or
|
|
deactivate asc.
|
|
|
|
Upon reading current asc status is returned.
|
|
|
|
What: /sys/bus/iio/devices/iio:deviceX/calibration_forced_value
|
|
Date: June 2020
|
|
KernelVersion: 5.8
|
|
Contact: linux-iio@vger.kernel.org
|
|
Description:
|
|
Contaminants build-up in the measurement chamber or optical
|
|
elements deterioration leads to sensor drift.
|
|
|
|
One can compensate for sensor drift by using forced
|
|
recalibration (frc). This is useful in case there's known
|
|
co2 reference available nearby the sensor.
|
|
|
|
Picking value from the range [400 1 2000] and writing it to the
|
|
sensor will set frc.
|
|
|
|
Upon reading current frc value is returned. Note that after
|
|
power cycling default value (i.e 400) is returned even though
|
|
internally sensor had recalibrated itself.
|